电感耦合等离子体串联质谱氨气模式测定土壤中的银
Determination of silver in soil by ICP tandem mass spectrometry ammonia mode
-
摘要: 本文建立了四酸溶样石墨消解-电感耦合等离子体串接质谱氨气反应模式测定土壤中银的方法.通过对氢气、氧气和氨气的3种模式下测定结果比较,选择了消除银干扰比较完全的氨气模式,同时测定银元素的107Ag和109Ag两个同位素以及土壤国家一级标准物质,结果证明方法准确可靠,方法检出限为0.003 mg·kg-1,精密度(RSD%)小于3%.本方法采用ICP-MS/MS技术,在线彻底去除多原子干扰,一个元素多同位素的测定具有数据双重验证效果.通过与分离富集处理样品以及在线干扰校正系数法测定银的方法相比,本方法节省了前处理时间同时具有更高的准确度.
-
关键词:
- 土壤 /
- 银 /
- 氨气模式 /
- 多原子离子 /
- 电感耦合等离子体串接质谱法
Abstract: A determination method of silver in soil by graphite digestion procedure combining ICP-MS/MS Ammonia mode was established. ICP-MS/MS hydrogen, oxygen and ammonia model measurement results were compared. The results showed that ammonia mode could completely remove polyatomic ion interferences. Two isotopes of silver were measured simultaneously and National standard substances were analyzed. The results proved that the method was accurate and reliable. The detection limit of the method was 0.003 mg·kg-1, and the precision was less than 3% (RSD%). This method used ICP-MS/MS technology completely to remove polyatomic interference online, and 107Ag and 109Ag isotope results could double verified the data. Compared with other works, this method saves the pre-processing time and has higher accuracy.-
Key words:
- soil /
- silver /
- ammonia mode /
- polyatomic ion /
- ICP-MS/MS
-
-
[1] 严冬,曾祥程,宋娟娥,等.电感耦合等离子体质谱测定水体中可溶性65元素[J].环境化学,2014,33(8):1418-1421. YAN D,ZENG C X,SONG J E,et al. Determination of soluble 65 elements in water by inductively coupled plasma mass spectrometry[J]. Environmental Chemistry, 2014,34(8):1418-1421(in Chinese).
[2] 邢智,漆亮.P507负载泡塑分离-电感耦合等离子体质谱法同时测定化探样品中银钨钼[J].岩矿测试,2014,33(4):486-490. XING Z,QI L. Simultaneous determination of Ag, W and Mo in geochemical exploration samples by ICP-MS Using P507 loaded foam for separation[J].Rock and Mineral Analysis, 2014,33(4):486-490(in Chinese).
[3] 乔波, 李中, 赵新建.活性炭吸附ICP-MS同时测定地球化学样品中的金和银[J].光谱实验室,2012,29(6):3721-3723. QIAO B,LI Z,ZHAO X J. Simultaneous determination of gold and sliver in geochemical samples by ICP-MS with adsorption of activated carbon[J]. Chinese Journal of Spectroscopy Laboratory,2012,29(6):3721-3723(in Chinese).
[4] 朱志刚,李美丽,孙元芳,等.ICP-MS测定银的干扰现象分析与方法建立[J].分析仪器,2016(5):70-74. ZHU Z G,LI M L,SUN Y F, et al. Analysis of interference phenomenon for determination of silver by ICP-MS[J]. Journal of Analytical Instruments, 2016 (5) 70-74(in Chinese).
[5] Agilent 8800 Triple Quadrupole ICP-MS:Understanding oxygen reaction mode in ICP-MS/MS[C]. Agilent Technologies, Inc., 2012(Published in December 20, 2012,Publication number:5991-1708EN). -

计量
- 文章访问数: 1037
- HTML全文浏览数: 1037
- PDF下载数: 43
- 施引文献: 0